Agilent DNA 7500 and
Agilent Technologies
Setting up the Assay Equipment and
Bioanalyzer
Before beginning the chip preparation protocol, ensure that the chip priming
station and the bioanalyzer are set up and ready to use.
You have t o
• replace the syringe at the chip priming station with each new DNA kit
• adjust the base plate of the chip priming station
• adjust the syringe clip at the chip priming station
• adjust the bioanalyzer’s chip selector
• set up the vortex mixer
• Finally, make sure that you start the software before you load the chip.
The Agilent DNA 7500 and DNA 12000 assay is a high sensitivity assay. Please read this
guide carefully and follow all instructions to guarantee satisfactory results.